Inadvertent temporary pacemaker lead placement in aortic sinus
Naveen Garg, Nagaraja Moorthy
Department of Cardiology, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow, Uttar Pradesh, India
Correspondence Address:
Nagaraja Moorthy Department of Cardiology, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow - 226 014, Uttar Pradesh India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/1995-705X.126885
|
|
Inadvertent placement of pacing leads into abnormal locations is potentially very dangerous. However, in emergency situations and without fluoroscopic guidance, these complications do occur. We report a case of an elderly male who underwent temporary pacemaker lead implantation without fluoroscopic guidance for cardiac arrest, but later the pacemaker lead was found to be in the non-coronary aortic sinus, but still capturing the myocardium. Interestingly, the post-pacing electrocardiography was mimicking atrial pacing. |
|
|
|